Evaluation of dental caries detection with quantitative light-induced fluorescence in comparison to different field of view devices.
Ontology highlight
ABSTRACT: This study evaluated dental caries detection ability between the Qraycam and Qraypen on the same dental caries lesions. A total of 178teeth from 61patients were imaged using Qraypen C®(QC) and Qraycam Pro®(QP) devices and evaluated using analysis software (QA2). Occlusal, secondary, and proximal dental caries were evaluated and scored according to International Caries Detection and Assessment System(ICDAS II) and X-ray criteria. Bland-Altman plots were used to compare quantitative light-induced fluorescence(QLF) parameters obtained from the different QLF devices. Sensitivity, specificity, and area under the receiver operating characteristic curve(AUROC) were calculated. The ΔFaver. of the QLF-parameters showed that the mean difference between the two different QLF devices was cl
SUBMITTER: Oh SH
PROVIDER: S-EPMC9005513 | biostudies-literature | 2022 Apr
REPOSITORIES: biostudies-literature