Evaluation of rapid antigen detection kits for detection of SARS-CoV-2 B.1.617.2 virus.
Ontology highlight
ABSTRACT: Aim: Currently, there is lack of data regarding rapid antigen detection (RAD) kits to detect SARS-CoV-2 B.1.617.2 virus. Objective: The purpose of this evaluation is to assess analytical sensitivity of 12 RAD kits against SARS-CoV-2 B.1.617.2. Study design: Analytical sensitivity was determined by limit of detection (LOD). A serial tenfold dilution set from a respiratory specimen collected from a COVID-19 patient infected by SARS-CoV-2 B.1.617.2 was used. RT-PCR was used as a reference method. Results: The LOD results showed that 11 and one RAD kits were 100- and 1000-fold less sensitive than RT-PCR respectively. Conclusion: The results showed that the RAD kits evaluated in this study may be used for first-line screening of the SARS-CoV-2 B.1.617.2 variant.