Validation of the Application of Solid Contact Ion-Selective Electrode for Off-Body Sweat Ion Monitoring.


ABSTRACT: The solid contact ion-selective electrode (ISE) is a promising skin-interfaced monitoring system for sweat ions. Despite a growing number of on-body usages of ISE with fancy new materials and device fabrications, there are very few reports attempting to validate ISE results with a gold standard technique. For this purpose, this work uses inductively coupled plasma-optical emission spectrometry (ICP-OES) as a reference technique to conduct a direct evaluation of the sweat sodium and potassium ion levels obtained by ISE in an off-body approach. Eight healthy male subjects were recruited to collect exercise-induced sweat. It was found that sweat sodium and potassium ions present a rather wide concentration range. The sweat sodium concentration did not vary greatly in an exercise period of hal

SUBMITTER: Liu H 

PROVIDER: S-EPMC9026306 | biostudies-literature | 2022 Apr