ABSTRACT: Background
Early detection and timely prophylaxis can retard the progression of osteoporosis. The purpose of this study was to determine the validity of peripheral Dual Energy X-ray Absorptiometry (DXA) test for osteoporosis screening. We examined peripheral bone mineral density (BMD) using AKDX-09 W-I DXA densitometer. Firstly, we acquired BMD data from manufacturer-supplied density-gradient phantoms and 30 volunteers to investigate its accuracy and precision, then we measured BMD for 150 volunteers using both AKDX (left forearm) and Hologic Discovery Wi (left forearm, left hip and L1 - L4 vertebrae) simultaneously. Correlation relationship of BMD results acquired from two instruments was assessed by simple linear regression analysis, the Receiver Operating Characteristic (ROC) cu