Unknown

Dataset Information

Comparison of early migration patterns between a ceramic and polyethylene liner in uncemented Trabecular Titanium cups: a 2-year randomized controlled trial of 52 hips using radiostereometric analysis.


ABSTRACT:

Background and purpose

Ceramic liners may reduce early stability of uncemented acetabular components due to higher stiffness. However, the bone ingrowth capacities of porous trabecular titanium might compensate for this effect. This prospective randomized trial quantifies migration patterns of the Delta-TT cup, and compares polyethylene and ceramic liners.

Patients and methods

Patients undergoing primary uncemented total hip arthroplasty with the Delta-TT cup and femoral stem with ceramic head were randomized to a polyethylene (n = 25) or ceramic (n = 28) liner. Radiostereometric analysis (RSA) radiographs, patient-reported hip function (HOOS-PS, OHS), and quality of life (EQ5D) were collected at baseline and 1.5, 3, 6, 12, and 24 months postoperatively. Model-based RSA was

SUBMITTER: Klaassen AD 

PROVIDER: S-EPMC9052219 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature