A ruthenium oxide and iridium oxide coated titanium electrode for pH measurement.


ABSTRACT: This paper reports the pH sensing capability of a ruthenium oxide (RuO2) and iridium oxide (IrO2) coated titanium (ROIOT) electrode. The characterization results indicated that the ROIOT electrode had a cracked morphology. The RuO2 and IrO2 particles were decorated on the surface of the electrode. The ROIOT electrode showed near-Nernstian sensitivity of -50.8 mV pH-1, with a wide detection range of pH 2-12. The response time was 4.0-13.5 s, which was fast and very sensitive to the pH change. The ROIOT electrode also demonstrated great detection reversibility and stability in various pH conditions. In the long-term experiment of 30 d, potential measurements using the ROIOT electrode had a minor fluctuation of 1.5 mV d-1. The