Development of an on-chip sample injection system with a 6-port valve incorporated in a microchip.
Ontology highlight
ABSTRACT: Micro-flow-injection analysis (μFIA) is amenable to high-throughput systems with lower consumption of sample and reagent volumes. On-chip sample injection methods are important to prevent reduced analytical performance associated with dead volumes and diffusion of sample solutions. In this study, we have developed an on-chip sample injection system with a small-sized 6-port valve incorporated on a microchip. The valve is made with a 3D printer and is a simple structure that can be easily operated manually. A sample solution in a loading channel can be injected by switching the valve from the load to injection position. Sample injection tests using resorufin solutions revealed that samples can be injected below 100 μL min-1, and the performance of the sample injection system is c