Dental imaging using an ultra-high resolution photon-counting CT system.
Ontology highlight
ABSTRACT: Clinical photon-counting CT (PCCT) offers a spatial resolution of about 200 µm and might allow for acquisitions close to conventional dental CBCTs. In this study, the capabilities of this new system in comparison to dental CBCTs shall be evaluated. All 8 apical osteolysis identified in CBCT were identified by both readers in all three PCCT scan protocols. Mean visibility scores showed statistical significant differences for root canals(p = 0.0001), periodontal space(p = 0.0090), cortical(p = 0.0003) and spongious bone(p = 0.0293) in favor of high and medium dose PCCT acquisitions. Overall, both devices showed excellent image quality of all structures assessed. Interrater-agreement showed high values for all protocols in all structures. Bland-Altman plots revealed a high concordance of both