Ontology highlight
ABSTRACT: Purpose
We aimed to evaluate the accuracy of T 1 and T 2 mappings derived from a multispectral pulse sequence (magnetic resonance image compilation, MAGiC®) on 1.5-T MRI and with conventional sequences [gradient echo with variable flip angle (GRE-VFA) and multi-echo spin echo (ME-SE)] compared to the reference values for the purpose of radiotherapy treatment planning.Methods
The accuracy of T 1 and T 2 measurements was evaluated with 2 coils [head and neck unit (HNU) and BODY coils] on phantoms using descriptive statistics and Bland-Altman analysis. The reproducibility and repeatability of T 1 and T 2 measurements were performed on 15 sessions with the HNU
SUBMITTER: Gouel P
PROVIDER: S-EPMC9065558 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature