Measuring glucose at the site of insulin delivery with a redox-mediated sensor.
Ontology highlight
ABSTRACT: Automated insulin delivery systems for people with type 1 diabetes rely on an accurate subcutaneous glucose sensor and an infusion cannula that delivers insulin in response to measured glucose. Integrating the sensor with the infusion cannula would provide substantial benefit by reducing the number of devices inserted into subcutaneous tissue. We describe the sensor chemistry and a calibration algorithm to minimize impact of insulin delivery artifacts in a new glucose sensing cannula. Seven people with type 1 diabetes undergoing automated insulin delivery used two sensing cannulae whereby one delivered a rapidly-acting insulin analog and the other delivered a control phosphate buffered saline (PBS) solution with no insulin. While there was a small artifact in both conditions that increased
SUBMITTER: Jacobs PG
PROVIDER: S-EPMC9067318 | biostudies-literature | 2020 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA