A new strategy for the sensitive electrochemical determination of nitrophenol isomers using β-cyclodextrin derivative-functionalized silicon carbide.
Ontology highlight
ABSTRACT: In the present study, thiol β-cyclodextrin (SH-CD) and ethylenediamine β-cyclodextrin (NH2-β-CD) were simultaneously grafted on the same interface of an Au NP deposited carboxyl SiC (Au@CSiC) nanocomposite. An electrochemical sensor for the simultaneous determination of nitrophenol isomers (o-nitrophenol, o-NP; p-nitrophenol, p-NP) using SH-CD and NH2-β-CD functionalized Au@SiC (Au@CSiC-SH/NH2-CD) nanocomposite was successfully constructed. Differential pulse voltammetry was used to quantify o-NP and p-NP within the concentration range of 0.01-150 μM under the optimal conditions. The detection limit (S/N = 3) of the sensor was 0.019 and 0.023 μM for o-NP and p-NP, respectively, indicating a low detecti
SUBMITTER: Wu S
PROVIDER: S-EPMC9076973 | biostudies-literature | 2018 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA