Unknown

Dataset Information

Technical note: Temperature and concentration dependence of water diffusion in polyvinylpyrrolidone solutions.


ABSTRACT:

Objective

The goal of this work is to provide temperature and concentration calibration of water diffusivity in polyvinylpyrrolidone (PVP) solutions used in phantoms to assess system bias and linearity in apparent diffusion coefficient (ADC) measurements.

Method

ADC measurements were performed for 40 kDa (K40) PVP of six concentrations (0%, 10%, 20%, 30%, 40%, and 50% by weight) at three temperatures (19.5°C, 22.5°C, and 26.4°C), with internal phantom temperature monitored by optical thermometer (±0.2°C). To achieve ADC measurement and fit accuracy of better than 0.5%, three orthogonal diffusion gradients were calibrated using known water diffusivity at 0°C and system gradient nonlinearity maps. Noise-floor fit bias was also controlled by limiting the maximum b-value used fo

SUBMITTER: Amouzandeh G 

PROVIDER: S-EPMC9090959 | biostudies-literature | 2022 May

REPOSITORIES: biostudies-literature