Fully implantable wireless batteryless vascular electronics with printed soft sensors for multiplex sensing of hemodynamics.
Ontology highlight
ABSTRACT: The continuous monitoring of hemodynamics attainable with wireless implantable devices would improve the treatment of vascular diseases. However, demanding requirements of size, wireless operation, and compatibility with endovascular procedures have limited the development of vascular electronics. Here, we report an implantable, wireless vascular electronic system, consisting of a multimaterial inductive stent and printed soft sensors capable of real-time monitoring of arterial pressure, pulse rate, and flow without batteries or circuits. Developments in stent design achieve an enhanced wireless platform while matching conventional stent mechanics. The fully printed pressure sensors demonstrate fast response times, high durability, and sensing at small bending radii. The device is monitore
SUBMITTER: Herbert R
PROVIDER: S-EPMC9094660 | biostudies-literature | 2022 May
REPOSITORIES: biostudies-literature