A fiber optic photoacoustic sensor for real-time heparin monitoring.


ABSTRACT: Heparin is a common anticoagulant, but heparin overdose is a common intensive care unit (ICU) medication error due to the narrow therapeutic window of heparin. Conventional methods to monitoring heparin suffer from long turnaround time, the need for skilled personnel, and low frequency of sampling. To overcome these issues, we describe here a fiber optic photoacoustic (PA) sensor for real-time heparin monitoring. The proposed sensor was validated with in vitro testing and in a simulated in vivo model using the following samples: (1) phosphate-buffered saline (PBS), (2) spiked human plasma, (3) spiked whole human blood, and (4) clinical samples from patients treated with heparin. Samples were validated by comparing the PA signal to the activated partial thromboplastin time (aPTT) as well as