Ontology highlight
ABSTRACT: Purpose
To compare cone mosaic metrics derived from adaptive optics scanning light ophthalmoscopy (AOSLO) images with those derived from Heidelberg Engineering SPECTRALIS High Magnification Module (HMM) images.Methods
Participants with contiguous cone mosaics had HMM imaging performed at locations superior and temporal to the fovea. These images were registered and averaged offline and then aligned to split-detection AOSLO images; 200 × 200-µm regions of interest were extracted from both modalities. Cones were semi-automatically identified by two graders to provide estimates of cone density and spacing.Results
Thirty participants with contiguous cone mosaics were imaged (10 males, 20 females; age range, 11-67 years). Image quality varied, and 80% of our participants
SUBMITTER: Wynne N
PROVIDER: S-EPMC9123519 | biostudies-literature | 2022 May
REPOSITORIES: biostudies-literature