Comparison Between the 24-hour Holter Test and 72-hour Single-Lead Electrocardiogram Monitoring With an Adhesive Patch-Type Device for Atrial Fibrillation Detection: Prospective Cohort Study.


ABSTRACT:

Background

There is insufficient evidence for the use of single-lead electrocardiogram (ECG) monitoring with an adhesive patch-type device (APD) over an extended period compared to that of the 24-hour Holter test for atrial fibrillation (AF) detection.

Objective

In this paper, we aimed to compare AF detection by the 24-hour Holter test and 72-hour single-lead ECG monitoring using an APD among patients with AF.

Methods

This was a prospective, single-center cohort study. A total of 210 patients with AF with clinical indications for the Holter test at cardiology outpatient clinics were enrolled in the study. The study participants were equipped with both the Holter device and APD for the first 24 hours. Subsequently, only the APD continued ECG monitoring for an additiona