Simplifying quantitative measurement of free radical species using an X-band EPR spectrometer.


ABSTRACT: The quantitative measurement of free radicals in liquid using an X-band electron paramagnetic resonance (EPR) was systematized. Quantification of free radicals by EPR requires a standard sample that contains a known spin amount/concentration. When satisfactory reproducibility of the sample material, volume, shape, and positioning in the cavity for EPR measurements can be guaranteed, a sample tested and a standard can be directly compared and the process of quantification can be simplified. The purpose of this study was to simplify manual quantitative EPR measurement. A suitable sample volume for achieving a stable EPR intensity was estimated. The effects of different solvents on the EPR sensitivity were compared. The stability and reproducibility of the EPR intensity of standard nitroxyl r