Development of sandwich electrochemiluminescence immunosensor for COVID-19 diagnosis by SARS-CoV-2 spike protein detection based on Au@BSA-luminol nanocomposites.
Ontology highlight
ABSTRACT: Coronavirus disease (COVID-19) is a new and highly contagious disease posing a threat to global public health and wreaking havoc around the world. It's caused by the Coronavirus that causes severe acute respiratory syndrome (SARS-CoV-2). In the current pandemic situation, rapid and accurate SARS-CoV-2 diagnosis on a large scale is critical for early-stage diagnosis. Early detection and monitoring of viral infections can aid in controlling and preventing infection in large groups of people. Accordingly, we developed a sensitive and high-throughput sandwich electrochemiluminescence immunosensor based on antigen detection for COVID-19 diagnosis (the spike protein of SARS-CoV-2). For the spike protein of SARS-CoV-2, the ECL biosensor had a linear range of 10 ng mL-1 to 10 µg mL
SUBMITTER: Hosseini M
PROVIDER: S-EPMC9130314 | biostudies-literature | 2022 Oct
REPOSITORIES: biostudies-literature