Ontology highlight
ABSTRACT: Background
Accurate cup placement during total hip arthroplasty (THA) is difficult because the intraoperative pelvic position changes even in supine patient position. We developed a device known as HipPointer; it corrects pelvic rotation and creates a functional pelvic plane as a reference. The aim of this study was to determine the device placement accuracy and investigate causes of error.Material and methods
HipPointer was used for cup placement in 353 hips of 308 patients who underwent direct-anterior-approach THA in supine position. The mean age at surgery and body mass index were 63.9 (17-90) years and 24.9 (16.6-42.0) kg/m2, respectively. The mean observation period was 40.5 (12-73) months. To investigate the accuracy of HipPointer, preoperative planning an
SUBMITTER: Nakasone S
PROVIDER: S-EPMC9133511 | biostudies-literature | 2022 Aug
REPOSITORIES: biostudies-literature