All-in-One Digital Microfluidics System for Molecular Diagnosis with Loop-Mediated Isothermal Amplification.
Ontology highlight
ABSTRACT: In this study, an "all-in-one" digital microfluidics (DMF) system was developed for automatic and rapid molecular diagnosis and integrated with magnetic bead-based nucleic acid extraction, loop-mediated isothermal amplification (LAMP), and real-time optical signal monitoring. First, we performed on- and off-chip comparison experiments for the magnetic bead nucleic acid extraction module and LAMP amplification function. The extraction efficiency for the on-chip test was comparable to that of conventional off-chip methods. The processing time for the automatic on-chip workflow was only 23 min, which was less than that of the conventional methods of 28 min 45 s. Meanwhile, the number of samples used in on-chip experiments was significantly smaller than that used in off-chip experiments; only
SUBMITTER: Hu S
PROVIDER: S-EPMC9138765 | biostudies-literature | 2022 May