Ontology highlight
ABSTRACT: Purpose
Cardiac pacing devices can detect and monitor atrial tachyarrhythmias (ATA) which increase the risk of thromboembolic complications. The aim of this study was to compare (1) two different atrial leads and (2) standard and optimized settings to detect ATA and reject far-field R-wave signal (FFRW).Methods
This was a prospective, randomized multi-center trial comparing St. Jude Medical OptiSense lead (tip-to-ring spacing 1.1 mm) and Tendril lead (tip-to-ring spacing 10.0 mm), having programmed atrial sensitivity at 0.2 mV and post-ventricular atrial blanking at 60 ms. We measured intra-atrial amplitudes of FFRW, intrinsic atrial signals, the amount of FFRW oversensing, and other inappropriate mode switching.Results
One hundred and ten patients were enrolled. Th
SUBMITTER: Pakarinen S
PROVIDER: S-EPMC9151574 | biostudies-literature | 2022 Apr
REPOSITORIES: biostudies-literature