ABSTRACT: Background
Cardiac resynchronization therapy (CRT) demands high energy utilization due to continuous biventricular pacing. Current technology allows 6 pacing configurations for a bipolar left ventricular (LV) lead. Understanding the energy requirements for each configuration will allow optimization of pacing output.Methods and results
Pacing impedance, LV voltage threshold at 1.5 ms (rheobase) and 0.4 ms and chronaxie were obtained in 6 LV configurations in 49 consecutive patients undergoing CRT implantation or replacement. Strength-duration curves were derived using the Lapicque formula. Pacing impedances and voltage thresholds at 1.5 and 0.4 ms, calculated minimum threshold energy at chronaxie, current drain, energy thresholds at 0.4 ms, and strength-duration curves were