Innovative Guide Extension Catheter Trapping Technique to Retrieve a Lost Stent From a Coronary Artery.


ABSTRACT: A stent that was being implanted in the left circumflex artery, to treat an iatrogenic dissection, became dislodged at the ostial left circumflex artery on a previously deployed stent implanted for the treatment of a distal left main bifurcation stenosis. We describe here a novel technique to retrieve the device safely. (Level of Difficulty: Advanced.).

SUBMITTER: Righetti S 

PROVIDER: S-EPMC9175196 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Innovative Guide Extension Catheter Trapping Technique to Retrieve a Lost Stent From a Coronary Artery.

Righetti Stefano S   Tresoldi Simone S   Calchera Ivan I   Alvarenga Carlos Enrique CE   Vandoni Pietro P  

JACC. Case reports 20220406 7


A stent that was being implanted in the left circumflex artery, to treat an iatrogenic dissection, became dislodged at the ostial left circumflex artery on a previously deployed stent implanted for the treatment of a distal left main bifurcation stenosis. We describe here a novel technique to retrieve the device safely. (<b>Level of Difficulty: Advanced.</b>). ...[more]