Three-year monitoring and comparison of results from two identical blood gas analyzers.


ABSTRACT:

Objectives

Measurement comparability between blood gas analyzers within a laboratory is of utmost importance. This study analyzed the data obtained from a three-year period.

Design and methods

For quality monitoring one blood sample was tested on two identical blood gas analyzers at each of three shifts/day for three years. Deming regression analysis was used to determine result correlation and statistical identity between the two analyzers for pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, and lactate. Failures in the two-analyzer comparison were determined by the performance limits from the Institute of Quality Management in Healthcare (IQMH) and from the manufacturer respectively.