Determination of the electronic portal imaging device pixel-sensitivity-map for quality assurance applications. Part 2: Photon beam dependence.


ABSTRACT:

Purpose

The EPID PSM is a useful EPID calibration method for QA applications. The dependence of the EPID PSM on the photon beam used to acquire it has been investigated in this study for the four available PSM methods. The aim is to inform upon the viability of applying a single PSM for all available photon beams to simplify PSM implementation and maintenance.

Methods

Four methods of PSM determination were each measured once in a single session on a single TrueBeam ® STx linac using 6 MV, 10 MV, 6 MV Flattening-Filter-Free (FFF), and 10 MV FFF photon beams. The resultant PSM was assessed for both intra- and inter-method beam dependence via comparison between PSM of the same method compared to the 6 MV PSM and via comparison between PSM of the same beam with the corresponding