Virtual non-contrast images calculated from dual-energy CT shoulder arthrography improve the detection of intraarticular loose bodies.


ABSTRACT:

Objective

This study aims to evaluate the image quality of virtual non-contrast (VNC) images calculated from dual-energy CT shoulder arthrography (DECT-A) and their ability to detect periosteal calcifications and intraarticular loose bodies.

Materials and methods

In 129 shoulders of 123 patients, DECT arthrography (80 kV/140 kV) was performed with diluted iodinated contrast material (80 mg/ml). VNC images were calculated with image postprocessing. VNC image quality (1 = worst, 5 = best), dose parameters, and CT numbers (intraarticular iodine, muscle, VNC joint fluid density) were assessed. Image contrast (iodine/muscle) and percentage of iodine removal were calculated. Two independent readers evaluated VNC and DECT-A images for periosteal calcifications and intraarticular lo