ABSTRACT: Introduction
Late gadolinium enhancement (LGE) cardiac magnetic resonance imaging (MRI) can be used to detect postablation atrial scar (PAAS) but its reproducibility and reliability in clinical scans across different magnetic flux densities and scar detection methods are unknown.Methods
Patients (n = 45) having undergone two consecutive MRIs (3 months apart) on 3T and 1.5T scanners were studied. We compared PAAS detection reproducibility using four methods of thresholding: simple thresholding, Otsu thresholding, 3.3 standard deviations (SD) above blood pool (BP) mean intensity, and image intensity ratio (IIR). We performed a texture study by dividing the left atrial wall intensity histogram into deciles and evaluated the correlation of the same decile of the two scans as we