Mechanical behavior of hybrid glenoid components compared to all-PE components: a finite element analysis.


ABSTRACT:

Purpose

The purpose of this finite element study was to compare bone and cement stresses and implant micromotions among all-polyethylene (PE) and hybrid glenoid components. The hypothesis was that, compared to all-PE components, hybrid components yield lower bone and cement stresses with smaller micromotions.

Methods

Implant micromotions and cement and bone stresses were compared among 4 all PE (U-PG, U-KG, A-KG, I-KG) and 2 hybrid (E-hCG, I-hPG) virtually implanted glenoid components. Glenohumeral joint reaction forces were applied at five loading regions (central, anterior, posterior, superior and inferior). Implant failure was assumed if glenoid micromotion exceeded 75 µm or cement stresses exceeded 4 MPa. The critical cement volume (CCV) was based on the percentage of ce