10-Year Impact of Transcatheter Aortic Valve Replacement Leaflet Design (Intra- Versus Supra-Annular) in Mortality and Hemodynamic Performance


ABSTRACT:

Background

The impact of transcatheter aortic valve replacement (TAVR) leaflet design on long-term device performance is still unknown. This study sought to compare the clinical and hemodynamic outcomes of intra- (IA) versus supra-annular (SA) TAVR designs up-to 10-years following implantation.

Methods

Consecutive patients with at least 5-years follow-up following TAVR for severe symptomatic aortic stenosis from June 2007 to December 2016 were included. Bioprosthetic valve failure (BVF) and hemodynamic valve deterioration (HVD) were defined according to VARC-3 updated definitions and estimated using cumulative incidence function to account for the competing risk of death.

Results

A total of 604 patients (82 years; 53% female) were analyzed and divided into IA (482) and SA (122) groups. Overall survival rates at 10-years were similar (IA 15%, 95%CI: 10–22; SA 11%, 95%CI: 6–20; p = 0.21). Compared to the SA TAVR, mean transaortic gradients were significantly higher and increased over time in the IA group. IA TAVRs showed higher 10-year cumulative incidences of BVF (IA 8% vs. SA 1%, p = 0.02) and severe HVD (IA 5% vs. SA 1%, p = 0.05). The occurrence of BVF and HVD in the IA group occurred primarily in the smallest TAVR devices (20–23-mm). After excluding these sizes, the cumulative incidences of BVF (IA 5% vs. SA 1%, p = 0.40) and severe HVD (IA 2% vs. SA 1%, p = 0.11) were similar.

Conclusion

In this study, TAVR leaflet design had no impact on survival at 10-years. IA devices showed higher transaortic gradients and cumulative incidences of HVD and BVF predominantly occurring in the smallest valve sizes. Graphical Abstract 10-year outcomes and hemodynamic performance of intra-annular vs. supra-annular TAVR. HVD, hemodynamic valve deterioration; TAVR, transcatheter aortic valve replacement; VARC, valve academic research consortium.

SUBMITTER: Scotti A 

PROVIDER: S-EPMC9215259 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC8649494 | biostudies-literature
| S-EPMC8639544 | biostudies-literature
| S-EPMC9186299 | biostudies-literature
| S-EPMC9091536 | biostudies-literature
| S-EPMC10493236 | biostudies-literature
| S-EPMC11839708 | biostudies-literature
| S-EPMC8311375 | biostudies-literature
| S-EPMC6475193 | biostudies-literature