Fast, Reusable, Cell Uniformly Distributed Membrane Filtration Device for Separation of Circulating Tumor Cells.
Ontology highlight
ABSTRACT: Isolation of circulating tumor cells (CTCs) is of great significance for the diagnosis, prognosis, and treatment of metastatic cancer. Among CTC capture methods independent of antibodies, membrane filtration-based methods have the advantages of simplicity, rapidity, and high throughput but usually have problems such as clogging, high pressure drop, and impaired cell viability. In this study, we designed and tested a reusable device that used horizontal rotor and fluid-assisted separation to capture CTCs by centrifugal membrane filtration, achieving simple, fast, highly efficient, and viable cell capture on traditional centrifuge. The average capture efficiency was 95.8% for different types of cancer cells with >90% survival, and the removal of white blood cells can reach 99.72% under four
SUBMITTER: Han J
PROVIDER: S-EPMC9219081 | biostudies-literature | 2022 Jun
REPOSITORIES: biostudies-literature
ACCESS DATA