A High-Throughput MEMS-Based Differential Scanning Calorimeter for Direct Thermal Characterization of Antibodies.


ABSTRACT: Calorimeters, which can be used for rapid thermal characterization of biomolecules, are getting intense attention in drug development. This paper presents a novel MEMS-based differential scanning calorimeter (DSC) for direct thermal characterization of protein samples. The DSC consisted of a pair of temperature sensors made by vanadium oxide (VOx) film with a temperature coefficient of resistivity of -0.025/K at 300 K, a microfluidic device with high thermal insulation (2.8 K/mW), and a Peltier heater for linear temperature scanning. The DSC exhibited high sensitivity (6.1 µV/µW), low noise (0.4 µW), high scanning rate (45 K/min), and low sample consumption volume (0.63 µL). The MEMS DSC was verified by measuring the temperature-induced denaturation of lysozyme at different pH,