Deviating from the Recommended Torque on Set Screws Can Reduce the Stability and Fatigue Life of Pedicle Screw Fixation Devices.


ABSTRACT: Background and Objectives: Using an appropriate torque to tighten set screws ensures the long-term stability of spinal posterior fixation devices. However, the recommended torque often varies between different devices and some devices do not state a recommended torque level. The purpose of this study is to evaluate the effect of set screw torque on the overall construct stability and fatigue life. Materials and Methods: Two commercial pedicle screw systems with different designs for the contact interface between the set screw and rod (Group A: plane contact, Group B: line contact) were assembled using torque wrenches provided with the devices to insert the set screws and tighten to the device specifications. The axial gipping capacity and dynamic mechanical stability of each

SUBMITTER: Wu LC 

PROVIDER: S-EPMC9228452 | biostudies-literature | 2022 Jun