Trends in maintenance status and usability of public automated external defibrillators during a 5-year on-site inspection.
Ontology highlight
ABSTRACT: This study aimed to assess the trend of the maintenance status and usability of public automated external defibrillators (AEDs). Public AEDs installed in Seoul from 2013 to 2017 were included. An inspector checked the maintenance status and usability of the AEDs annually using a checklist. During the study period, 23,619 AEDs were inspected. Access to the AEDs was improved, including the absence of obstacles near the AEDs (from 90.2% in 2013 to 99.1% in 2017, p < 0.0001) and increased AED signs (from 34.3% in 2013 to 91.3% in 2017, p < 0.0001). The rate of AEDs in normal operation (from 94.0% in 2013 to 97.6% in 2017, p < 0.0001), good battery status (from 95.6% in 2013 to 96.8% in 2017, p = 0.0016), and electrode availability increased (from 97.1% in 2013 to 99.0% in 2017, p < 0.0001); th