Computerized tomography image correlation of His bundle/deep septal pacing location and outcomes: an analysis from the Canberra HIs bundle/deep septal Pacing Study (CHIPS).


ABSTRACT:

Background

Localisation of the conduction system under fluoroscopy is not easy and the ideal location of the pacing leads in physiological pacing is still being debated.

Objective

The primary aim was to assess the lead locations using cardiac CT scan. Secondary aims were clinical outcomes including success and safety of the procedure and lead performance.

Methods

Of the 100 consecutive patients who received physiological pacing, 34 patients underwent follow-up cardiac CT scan. The four different types of pacing were identified as His bundle (HBP), para-Hisian, left bundle branch (LBBP), and deep septal pacing.

Results

Most patients had successful HBP via the right atrium (RA) (87.5%) as compared to the right ventricle (RV) (12.5%). Lower thresholds were observe