Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900.


ABSTRACT:

Background

Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR.

Methods

Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF.

Results

High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect.

Conclusion

Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics.

SUBMITTER: Ruge H 

PROVIDER: S-EPMC9238356 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Bioprosthetic Valve Fracturing: <i>In vitro</i> Testing of Edwards PERIMOUNT Model P 2900.

Ruge Hendrik H   Alvarez-Covarrubias Hector A HA   Deutsch Oliver O   Alalawi Zahra Z   Vitanova Keti K   Lange Rüdiger R  

Frontiers in cardiovascular medicine 20220602


<h4>Background</h4>Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR.<h4>Methods</h4>Using four Perimount 19- and 21-mm valves, <i>in-vitro</i> high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dila  ...[more]

Similar Datasets

| S-EPMC10100054 | biostudies-literature
| S-EPMC8811120 | biostudies-literature
| S-EPMC8940997 | biostudies-literature
| S-EPMC5577954 | biostudies-literature
| S-EPMC5037665 | biostudies-literature
| S-EPMC8167470 | biostudies-literature
| S-EPMC11830275 | biostudies-literature
| S-EPMC3783116 | biostudies-literature
| S-EPMC8514018 | biostudies-literature
| S-EPMC10443654 | biostudies-literature