Evaluation of Sibel's Advanced Neonatal Epidermal (ANNE) wireless continuous physiological monitor in Nairobi, Kenya.


ABSTRACT:

Background

Neonatal multiparameter continuous physiological monitoring (MCPM) technologies assist with early detection of preventable and treatable causes of neonatal mortality. Evaluating accuracy of novel MCPM technologies is critical for their appropriate use and adoption.

Methods

We prospectively compared the accuracy of Sibel's Advanced Neonatal Epidermal (ANNE) technology with Masimo's Rad-97 pulse CO-oximeter with capnography and Spengler's Tempo Easy reference technologies during four evaluation rounds. We compared accuracy of heart rate (HR), respiratory rate (RR), oxygen saturation (SpO2), and skin temperature using Bland-Altman plots and root-mean-square deviation analyses (RMSD). Sibel's ANNE algorithms were optimized between each round. We created Clarke error g