Reduced motion external defibrillation: Reduced subject motion with equivalent defibrillation efficiency validated in swine.


ABSTRACT:

Background

External defibrillators are used for arrhythmia cardioversion and for defibrillating during cardiac arrest. During defibrillation, short-duration biphasic pulses cause intense motion due to rapid chest-wall muscle contraction. A reduced motion external defibrillator (RMD) was constructed by integrating a commercial defibrillator with a Tetanizing-waveform generator. A long-duration, low-amplitude, tetanizing waveform slowly stimulated the chest musculature before the biphasic pulse, reducing muscle contraction during the shock.

Objective

The purpose of this study was to evaluate RMD defibrillation in swine for subject motion during defibrillation pulses and for defibrillation effectiveness. RMD defibrillation can reduce the duration of arrhythmia ablation therapy