A force measurement platform for a vitreoretinal surgical simulator using an artificial eye module integrated with a quartz crystal resonator.
Ontology highlight
ABSTRACT: To provide quantitative feedback on surgical progress to ophthalmologists practicing inner limiting membrane (ILM) peeling, we developed an artificial eye module comprising a quartz crystal resonator (QCR) force sensor and a strain body that serves as a uniform force transmitter beneath a retinal model. Although a sufficiently large initial force must be loaded onto the QCR force sensor assembly to achieve stable contact with the strain body, the highly sensitive and wide dynamic-range property of this sensor enables the eye module to detect the slight forceps contact force. A parallel-plate strain body is used to achieve a uniform force sensitivity over the 4-mm-diameter ILM peeling region. Combining these two components allowed for a measurable force range of 0.22 mN to 29.6 N with a sen