Unknown

Dataset Information

0

A handheld intelligent single-molecule binary bioelectronic system for fast and reliable immunometric point-of-care testing.


ABSTRACT: Molecular tests are highly reliable and sensitive but lack portability and are not simple to use; conversely, easy-to-use antigenic tests still lack high performance. BioScreen combines single-molecule sensitivity and outstanding reliability with ultraportability and simplicity of use. This digital platform is capable of artificial intelligence-based binary classification at the limit of identification of a single marker/virus in 0.1 ml. The diagnostic sensitivity, specificity, and accuracy reach 99.2% as validated through 240 assays, including a pilot clinical trial. The versatile immunometric system can detect the SARS-CoV-2 virus, spike S1, and immunoglobulin G antigen proteins in saliva, blood serum, and swab. BioScreen has a small footprint comprising a disposable cartridge and a handheld electronic reader connected to a smart device. The sample handling is minimal, and the assay time to result is 21 min. Reliable and sensitive self-testing with an ultraportable and easy-to-use diagnostic system operated directly by a patient holds the potential to revolutionize point-of-care testing and early diagnosis.

SUBMITTER: Macchia E 

PROVIDER: S-EPMC9258948 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

A handheld intelligent single-molecule binary bioelectronic system for fast and reliable immunometric point-of-care testing.

Macchia Eleonora E   Kovács-Vajna Zsolt M ZM   Loconsole Daniela D   Sarcina Lucia L   Redolfi Massimiliano M   Chironna Maria M   Torricelli Fabrizio F   Torsi Luisa L  

Science advances 20220706 27


Molecular tests are highly reliable and sensitive but lack portability and are not simple to use; conversely, easy-to-use antigenic tests still lack high performance. BioScreen combines single-molecule sensitivity and outstanding reliability with ultraportability and simplicity of use. This digital platform is capable of artificial intelligence-based binary classification at the limit of identification of a single marker/virus in 0.1 ml. The diagnostic sensitivity, specificity, and accuracy reac  ...[more]