Evidence to support magnetic resonance conditional labelling of all pacemaker and defibrillator leads in patients with cardiac implantable electronic devices.


ABSTRACT:

Aims

Many cardiac pacemakers and defibrillators are not approved by regulators for magnetic resonance imaging (MRI). Even following generator exchange to an approved magnetic resonance (MR)-conditional model, many systems remain classified 'non-MR conditional' due to the leads. This classification makes patient access to MRI challenging, but there is no evidence of increased clinical risk. We compared the effect of MRI on non-MR conditional and MR-conditional pacemaker and defibrillator leads.

Methods and results

Patients undergoing clinical 1.5T MRI with pacemakers and defibrillators in three centres over 5 years were included. Magnetic resonance imaging protocols were similar for MR-conditional and non-MR conditional systems. Devices were interrogated pre- and immediately

SUBMITTER: Bhuva AN 