Unknown

Dataset Information

Practical Considerations for Clinical Laboratories in Top-down Approach for Assessing the Measurement Uncertainty of Clinical Chemistry Analytes.


ABSTRACT:

Background

The top-down (TD) approach using internal quality control (IQC) data is regarded a practical method for estimating measurement uncertainty (MU) in clinical laboratories. We estimated the MU of 14 clinical chemistry analytes using the TD approach and evaluated the effect of lot changes on the MU.

Methods

MU values were estimated using subgrouping by reagent lot changes or using the data as a whole, and both methods were compared. Reagent lot change was simulated using randomly generated data, and the mean values and MU for two IQC datasets (different QC material lots) were compared using statistical methods.

Results

All MU values calculated using subgrouping were lower than the total values; however, the average differences were minimal. The simulation showe

SUBMITTER: Gu H 

PROVIDER: S-EPMC9277045 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature