ABSTRACT: Purpose
Accurate two-dimensional (2D) profile measurements at submillimeter precision are necessary for proton beam commissioning and periodic quality assurance (QA) purposes and are currently performed at our institution with a commercial scintillation detector (Lynx PT) with limited means for independent checks. The purpose of this work was to create an independent dosimetry system consisting of an in-house optical scanner and a BaFBrI:Eu2+ storage phosphor dosimeter by: (a) determining the optimal settings for the optical scanner, (b) measuring 2D proton spot profiles with the storage phosphors, and (c) comparing them to similar measurements using a commercial scintillation detector.Methods