Characterization of positioning uncertainties in PET-CT-MR trimodality solutions for radiotherapy.


ABSTRACT: The purpose of this study was to evaluate the positioning uncertainties of two PET/CT-MR imaging setups, C1 and C2. Because the PET/CT data were acquired on the same hybrid device with automatic image registration, experiments were conducted using CT-MRI data. In C1, a transfer table was used, which allowed the patient to move from one imager to another while maintaining the same position. In C2, the patient stood up and was positioned in the same radiotherapy treatment position on each imager. The two setups provided a set of PET/CT and MR images. The accuracy of the registration software was evaluated on the CT-MRI data of one patient using known translations and rotations of MRI data. The uncertainties on the two setups were estimated using a phantom and a cohort of 30 patients. The acc