Optimization and evaluation of a two-stage chromogenic assay procedure for measurement of emicizumab plasma levels.
Ontology highlight
ABSTRACT: Emicizumab mimics the hemostatic activity of activated factor VIII (FVIIIa) within the tenase complex. Despite functional similarities between FVIIIa and emicizumab, conventional laboratory methods designed for monitoring of FVIII activity are inappropriate for the measurement of emicizumab. At present, a modified one stage (FVIII) assay (mOSA) is mainly used for emicizumab monitoring. Two-stage chromogenic FVIII assays based on human factors can be used, although limited performance due to lack of corresponding optimization might be observed. Furthermore, the presence of FVIII or anticoagulants in the patient sample may falsify assay results. To address these issues, we optimized and evaluated a two-stage chromogenic assay (emi-tenase) for measurement of emicizumab in plasma samples. Heat