ABSTRACT: Background and purpose
The study evaluated the differences in leaf positioning deviations by the log files of three advanced accelerators with two delivery techniques, and established specific assessment parameters of leaf positioning deviations for different types of accelerators.Methods
A total of 420 treatment plans with 5 consecutive treatment log files were collected from the Trilogy, TrueBeam and Halcyon accelerators. Millennium MLC was equipped on the Trilogy and TrueBeam accelerators. A jawless design and dual-layer MLC were adopted on the Halcyon accelerator. 70 IMRT and 70 VMAT plans were selected randomly on each accelerator. The treatment sites of all plans included head and neck, chest, breast, pelvis and other sites. The parsing tasks for 2100 log files were p