ABSTRACT: Objectives
Resonance frequency analysis (RFA) is used to monitor implant stability. Its output, the Implant Stability Quotient (ISQ), supposedly correlates with insertion torque, a common measurement of primary stability. However, the reliability of RFA in condensed bone remains unclear.Material and methods
In this human cadaver study in edentulous jaws and fresh extraction sockets, implants were inserted using a split-mouth approach into condensed or untreated bone. Mean ISQ, peak insertion torque, and pre- and postoperative bone volume fractions (BV/TV) were assessed.Results
In edentulous jaws, insertion torque and ISQ correlated both in untreated (r = 0.63, p = 0.02) and in condensed (r = 0.82, p < 0.01) bone. In extraction sockets, insertion torque and ISQ only