Variation of Mean Absolute Relative Differences of Continuous Glucose Monitoring Systems Throughout the Day.


ABSTRACT:

Background

There is an increasing use of continuous glucose monitoring (CGM) by people with diabetes. Measurement performance is often characterized by the mean absolute relative difference (MARD). However, MARD is influenced by a number of factors and little is known about whether MARD is stable throughout the day.

Material and methods

A total of 24 participants with type 1 diabetes were enrolled in the study. The study was performed for seven in-patient days. Participants wore two CGM systems in parallel and performed additional frequent blood glucose (BG) measurements. On two days, glucose excursions were induced.MARD was calculated between pairs of CGM and BG values, with BG values serving as reference values. ARD values calculated from CGM-BG pairs were grouped by hour