Ultrasonic cleaning is effective in removing carbonized clots and tissue from the insulation-tipped diathermic knife-2.


ABSTRACT:

Objectives

Since carbonized clots and tissue (debris) tend to adhere firmly to the tip of the endoscopic submucosal dissection (ESD) knife as the procedure proceeds, manual removing the firm debris is often challenging and time-consuming. Recently, effective ultrasonic cleaning for other medical devices has been reported. The aim of the present study was to clarify whether ultrasonic cleaning is effective in removing the debris on the insulation-tipped diathermic (IT) knife-2.

Methods

This study was an ex-vivo experimental randomized study. A total of 40 IT knife-2 knives with debris on their tip surfaces were prepared and randomly assigned to two groups (Group A and Group B). The knives in Group A were cleaned using the conventional scrubbing method for 30 s (conventional c