Comparing the safety and effectiveness of dedicated radiofrequency transseptal wires to electrified metal guidewires.


ABSTRACT:

Background

Application of electrocautery to a metal guidewire is used by some operators to perform transseptal puncture (TSP). Commercially available dedicated radiofrequency (RF) guidewires may represent a better alternative. This study compares the safety and effectiveness of electrified guidewires to a dedicated RF wire.

Methods

TSP was performed on freshly excised porcine hearts using an electrified 0.014″ or 0.032″ guidewire under various power settings and was compared to TSP using a dedicated RF wire with 5 W power (0.035″ VersaCross RF System, Baylis Medical). The primary endpoint was the number of attempts required to achieve TSP. Secondary endpoints included the rate of TSP failure, TSP consistency, the effect of the distance between the tip of the guidewire and th