CLinical Assessment of WEB device in Ruptured aneurYSms (CLARYS): results of 1-month and 1-year assessment of rebleeding protection and clinical safety in a multicenter study.


ABSTRACT:

Background

The primary goal of the CLARYS study is to assess the protection against rebleeding when treating ruptured bifurcation aneurysms with the Woven EndoBridge (WEB) device.

Methods

The CLARYS study is a prospective, multicenter study conducted in 13 European centers. Patients with ruptured bifurcation aneurysms were consecutively included between February 2016 and September 2017. The primary endpoint was defined as the rebleeding rate of the target aneurysm treated with the WEB within 30 days postprocedure. Secondary endpoints included periprocedural and postprocedural adverse events, total procedure and fluoroscopy times, and modified Rankin Scale score at 1 month and 1 year.

Results

Sixty patients with 60 ruptured bifurcation aneurysms to be treated with the WEB were included. A WEB device was successfully implanted in 93.3%. The rebleeding rate at 1 month and 1 year was 0%. The mean fluoroscopy time was 27.0 min. Twenty-three periprocedural complications were observed in 18 patients and resolved without sequelae in 16 patients. Two of these complications were attributed to the procedure and/or the use of the WEB, leading to a procedure/device-related intraoperative complication rate of 3.3%. Overall mortality at 1 month and 1 year was 1.7% and 3.8%, respectively and overall morbidity at 1 month and 1 year was 15% and 9.6%, respectively. WEB-related 1-month and 1-year morbidity and mortality was 0%.

Conclusions

The interim results of CLARYS show that the endovascular treatment of ruptured bifurcation aneurysms with the WEB is safe and effective and, in particular, provides effective protection against rebleeding. It may induce profound change in the endovascular management of ruptured bifurcation aneurysms.

SUBMITTER: Spelle L 

PROVIDER: S-EPMC9304095 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

CLinical Assessment of WEB device in Ruptured aneurYSms (CLARYS): results of 1-month and 1-year assessment of rebleeding protection and clinical safety in a multicenter study.

Spelle Laurent L   Herbreteau Denis D   Caroff Jildaz J   Barreau Xavier X   Ferré Jean-Christophe JC   Fiehler Jens J   Januel Anne-Christine AC   Costalat Vincent V   Liebig Thomas T   Bourcier Romain R   Möhlenbruch Markus M   Berkefeld Joachim J   Weber Werner W   Mihalea Cristian C   Ikka Léon L   Ozanne Augustin A   Cognard Christophe C   Narata Ana Paula AP   Bibi Richard Edwige RE   Gauvrit Jean-Yves JY   Raoult Hélène H   Velasco Stéphane S   Buhk Jan-Hendrik JH   Chalumeau Vanessa V   Bester Maxim M   Desal Hubert H   du Mesnil de Rochemont Richard R   Bohner Georg G   Fischer Sebastian S   Biondi Alessandra A   Grimaldi Lamiae L   Moret Jacques J   Byrne James J   Pierot Laurent L  

Journal of neurointerventional surgery 20210907 8


<h4>Background</h4>The primary goal of the CLARYS study is to assess the protection against rebleeding when treating ruptured bifurcation aneurysms with the Woven EndoBridge (WEB) device.<h4>Methods</h4>The CLARYS study is a prospective, multicenter study conducted in 13 European centers. Patients with ruptured bifurcation aneurysms were consecutively included between February 2016 and September 2017. The primary endpoint was defined as the rebleeding rate of the target aneurysm treated with the  ...[more]

Similar Datasets

| S-EPMC10314057 | biostudies-literature
| S-EPMC7984684 | biostudies-literature
| S-EPMC8886787 | biostudies-literature
| S-EPMC11832578 | biostudies-literature
| S-EPMC8440913 | biostudies-literature
| S-EPMC9131500 | biostudies-literature
| S-EPMC10687667 | biostudies-literature
| S-EPMC9424161 | biostudies-literature
| S-EPMC11762049 | biostudies-literature
| S-EPMC7960317 | biostudies-literature